905 nm Raman DTS: engineering implementation and validation report
A complete engineering route from a 905 nm pulsed pump and approximately 870/943 nm Raman channels through dual APDs, synchronous acquisition, differential-loss correction, temperature inversion, calibration and prototype validation.
Updated 2026-11-24 · 21 min read
Raman distributed temperature sensing (DTS) uses the round-trip time of pulses in an optical fibre and the ratio of two Raman bands to reconstruct a temperature profile. A 905 nm pulsed source can support this architecture, but the project is not a ranging module connected to a fibre.
Pump spectral width and drift, the weak anti-Stokes channel near 870 nm, rejection of residual 905 nm light, differential attenuation between channels, detector overload and system calibration all affect temperature uncertainty. This report defines a validation-oriented prototype and keeps suggested targets separate from guaranteed product specifications.
Primary conclusion: Do not judge feasibility from the 905 nm laser peak power alone. The critical path is the long-fibre SNR of the anti-Stokes channel, short-wavelength fibre behaviour, pump rejection and dual-channel calibration.
1 Scope and design assumptions
The project objective is a single-ended, 905 nm pumped, dual-Raman-channel DTS prototype. It must produce a temperature profile along a fibre and demonstrate range, spatial resolution, temperature error, repeatability and stability in controlled temperature zones.
Required functions include:
- Controlled and measurable centre wavelength, pulse width, pulse energy, repetition rate and trigger timing.
- A common transmit/receive fibre port, 905 nm rejection and separation of the anti-Stokes and Stokes bands.
- Two independent low-noise receive channels with synchronous sampling and managed dynamic range.
- Time-to-distance mapping with fixed system-delay compensation.
- Channel-response correction, differential-loss correction and temperature inversion.
- Reference-temperature zones, uniform heated sections, temperature steps, long-run stability and fault-injection tests.
| Item | Current assumption | Still to freeze |
|---|---|---|
| Pump wavelength | 905 nm nominal; measure actual centre and drift | Wavelength bin and drift over current and temperature |
| Raman shift | Approximately 440 cm-1 for the initial estimate | Measured spectral width, passbands and detector response |
| Sensing fibre | Model not selected | Short-wavelength attenuation, mode structure, length, splice and bend losses |
| Range | Validate first at 0.5-1 km | Final 1, 5 or 10 km target and environmental conditions |
| Performance | Use 1-2 m and +/-1-2 °C as trial design inputs | Update time, spatial sampling, error definition and temperature range |
| Form factor | Benchtop prototype first | Size, power, interfaces, ingress protection and production cost |
Design rule: Every target must become a testable acceptance condition. Do not turn an unverified research target into a product guarantee.
2 Measurement principle and wavelengths
Each fibre position generates backscatter. The arrival time identifies position; the anti-Stokes/Stokes intensity ratio provides a temperature-sensitive measurand. The short-wavelength Raman spectrum has finite width and must be measured rather than inferred from nominal lines.
For a 905 nm pump:
- Pump wavenumber: approximately 11049.7 cm-1.
- With a 440 cm-1 silica Raman shift, the anti-Stokes channel is approximately 870.3 nm.
- The Stokes reference is approximately 942.5 nm.
| Channel | Relation | Approximate centre | Role |
|---|---|---|---|
| Anti-Stokes | v0 + delta-v | 870.3 nm | Temperature-sensitive and usually the weakest channel |
| Pump | v0 | 905 nm | Emitted light; residual leakage must be rejected |
| Stokes | v0 - delta-v | 942.5 nm | Weaker temperature dependence; reference channel |

A common single-ended model uses the dark-corrected, gain-corrected signals PAS(z) and PS(z), with R(z) = PAS(z) / PS(z):
T(z) = gamma / [C - delta-alpha * z - ln(R(z))]
Here gamma is derived from the Raman shift, C is a calibrated system constant and delta-alpha is the differential attenuation coefficient. The sign convention must match the calibration equation. Do not hard-code unverified constants. Estimate C, differential attenuation and section losses from known-temperature sections and controlled long-fibre tests.
Time, distance and resolution
z = c * t / (2 * ng)
delta-z approximately c * tau / (2 * ng)
where ng is the group index and tau is pulse width.
| Pulse width | Ideal resolution, ng approximately 1.468 | Engineering consequence |
|---|---|---|
| 5 ns | 0.51 m | Low energy; receiver bandwidth and noise matter more |
| 10 ns | 1.02 m | Good starting point for approximately 1 m resolution |
| 20 ns | 2.04 m | More single-pulse energy but coarser detail |
| 50 ns | 5.11 m | Useful for first long-fibre SNR validation |

3 System architecture
The recommended architecture is modular: pulsed 905 nm source; broadband common-path transmit/receive optics; 905 nm rejection and Raman wavelength separation; two APD/TIA channels; synchronous ADC/FPGA acquisition; and host software for temperature inversion and diagnostics.

| Subsystem | Input/output | Required function | Key validation |
|---|---|---|---|
| 905 nm transmitter | Trigger to optical pulse | Stable wavelength, controlled pulse/repetition, energy monitor | Pulse width, jitter, energy and optical waveform |
| Common-path optics | Pump to/from fibre | Low-loss coupling, transmit/return isolation, 870-943 nm path | Insertion loss, return loss, passbands and pump isolation |
| Spectral splitter | Mixed return to two bands | Reject 905 nm and separate anti-Stokes/Stokes | Passband, stopband and strong pump-leakage test |
| Receiver | Light to voltage | Low noise, low drift, suitable bandwidth, controlled overload recovery | Dark noise, linearity, overload and drift |
| Acquisition | Analog to digital | Synchronous sampling, averaging, common time base | Channel delay, clock, frame loss and window tests |
| Temperature software | Digital returns to profile | Filtering, distance mapping, gain and attenuation correction | Reference-temperature comparison and uncertainty report |
The common-path components must work not only at 905 nm but also at approximately 870 and 943 nm. Do not assume a 905 nm narrowband circulator or collimator remains adequate outside its stated band. Strong end-face or pump leakage can saturate the return path and widen the near-end dead zone. Monitor transmitted pulse energy so each shot can be normalised or flagged.
4 Suggested prototype targets
| Item | Initial target | Validation | Freeze condition |
|---|---|---|---|
| Centre wavelength | 905 nm nominal | Cold/hot spectral sweep at multiple currents | Allowed range and drift |
| Pulse width | Start near 10 ns; evaluate 5/10/20 ns | High-speed detector and oscilloscope | Resolution requirement |
| Pulse-energy stability | Initial 1-2% study, RMS or peak-to-peak stated | At least 10,000 pulses | Distinguish emitter output from in-fibre energy |
| Repetition rate | Programmable; a 5 km return window implies about 20.4 kHz upper bound | Long-window and trigger test | Working rate plus guard time |
| Spatial resolution | Target 1-2 m in the trial | Temperature step or short heated zone | Fully defined 10-90% method |
| Range | First 0.5-1 km | Long-fibre reference zone and SNR statistics | Link-budget result |
| Temperature error | Explore +/-1-2 °C | At least three reference temperatures | Statistics, range and confidence interval |
| Temperature resolution | Evaluate whether 0.2-0.5 °C is possible | Repeated measurements at stable temperature | Range, averaging and bandwidth |
| Update time | Configurable, for example 1-10 s | Record update rate and noise | Trade against averaging |
| Environment | Benchtop first | Ambient, vibration and supply changes | Product operating/storage conditions |
| Laser safety | Risk assessment of the complete product | Classification, labels and interlocks | Accessible-emission analysis |
Derived relations:
| Quantity | Relation | Design implication |
|---|---|---|
| Spatial resolution | delta-z approximately ctau/(2ng) | Shorter pulses improve ideal resolution but reduce energy |
| Maximum unaliased pulse rate | f_rep < c/(2ngL) | Longer fibre reduces the allowable repetition rate |
| Pulse energy | Ep approximately Ppeak*tau for a rectangle | Integrate the real optical waveform for design |
| Average power | Pavg = Ep*f_rep | Drives thermal, power and safety budgets |
| Distance map | z = c(t-t0)/(2ng) | Calibrate fixed delay t0 and group index ng |
5 905 nm source and pulsed driver
The 905 nm laser is a pump candidate. Its ranging datasheet is not automatically a DTS specification. Re-qualify the device at the selected pulse width, in-fibre energy, repetition rate and long-term operating point.
Required source characterisation includes:
- Wavelength versus case temperature, current and pulse condition.
- FWHM and spectral shape under pulsed drive.
- Rise, fall, overshoot, ringing and pulse-to-pulse jitter.
- Emitter energy and energy coupled into the sensing fibre.
- Long-run drift after warm-up.
- Peak current, duty cycle, thermal resistance, junction-temperature estimate and package stress.
| Characterisation item | Required measurement | Implementation note |
|---|---|---|
| Centre wavelength and drift | Peak wavelength versus case temperature, current and pulse condition | Build lambda-peak = f(Tcase, I, tau, f_rep) |
| Spectral width | FWHM, shape and pulse-related side modes | Do not substitute CW spectrum for pulsed spectrum |
| Pulse waveform | FWHM, edges, overshoot, ringing and shot-to-shot jitter | Use a high-speed photodetector and adequate scope |
| Output energy | Bare-emitter energy and in-fibre pulse energy | Use a calibrated energy measurement or waveform integration |
| Long-run drift | Wavelength, energy and pulse width after warm-up | Establish warm-up and health-monitor thresholds |
| Thermal/lifetime risk | Peak current, duty cycle, thermal resistance and junction temperature | Do not exceed forward-current, reverse-voltage or junction limits |
Driver blocks:
| Block | Function | Design note |
|---|---|---|
| Supply and energy storage | Low-noise bias plus transient pulse current | Low-ESL/ESR capacitor bank and small current loop |
| Pulse switch | Fast on/off with adjustable width and repetition | Check current, speed, capacitance and safe operating area |
| Current limit | Protect the laser and control peak current | Adequate sensing bandwidth and startup/fault limits |
| Trigger and clock | Internal/external trigger and synchronous output | Share or lock the ADC reference clock |
| Waveform monitor | Current, optical pulse and temperature | Current and optical waveforms are not identical |
| Protection | Open/short, over-temperature, interlock and rate limit | Fail-off by default and log fault codes |
Increasing pulse energy does not guarantee a proportional temperature improvement. Higher peak power can increase connector reflection, near-end saturation, nonlinearity or thermal load. Establish timing first with a pulse near 10 ns, then measure both Raman channels and dynamic range, and finally build the Pareto surface of accuracy, update time and range.
Before system integration, the source should have repeatable optical pulses, a defined in-fibre pulse energy, traceable wavelength and drift data, and synchronous monitoring of current and optical output.
6 Fibre link
Common telecom single-mode fibre is designed around 1310/1550 nm. ITU-T G.652 cable cutoff is specified at or below 1260 nm and its attenuation and mode properties should not be assumed unchanged at 870-943 nm. The 905 nm system must measure the actual candidate fibre at all three wavelengths.
| Fibre property | Test band/condition | Output | System impact |
|---|---|---|---|
| Attenuation | Approximately 870, 905 and 943 nm | alpha870, alpha905, alpha943 in dB/km | Range and differential-loss correction |
| Mode behaviour | Target band and installed length | Single/multimode, modal distribution and stability | Pulse broadening, return shape and repeatability |
| Bend sensitivity | Defined radius, turns and positions | Loss change and profile error | A bend can appear as a false temperature change |
| Splice and connector | Face quality, fusion and mating cycles | Insertion loss, reflection and channel steps | Local artefacts and dead zones |
| Temperature exposure | Intended range and thermal cycles | Attenuation and jacket/coating changes | Calibration and deployment reliability |
| Raman response | Candidate fibre under pump | Actual anti-Stokes/Stokes signal and SNR | Confirms the useful passbands |
The light path should cover the pump and both Raman returns. Check circulator/coupler isolation, insertion loss, return loss, polarisation-dependent loss and wavelength range. Measure 905 nm rejection under real optical power; a nominal optical density is not a system-level guarantee.
The link-budget template must include in-fibre pulse energy, fibre length, round-trip attenuation, splice/connector loss, Raman collection efficiency, detector noise, filter transmission and received SNR at the far end under worst-case temperature and loss.
| Budget item | Value to record | Note |
|---|---|---|
| In-fibre pulse energy | E_launch = ____ nJ/uJ | Measure at the sensing-fibre input, not at the chip |
| Fibre length | L = ____ km | Record both reel and deployed length |
| Round-trip loss | Approximately 2alphaL, separately for anti-Stokes and Stokes | Add connector, splice and circulator/coupler loss |
| Raman collection efficiency | Measured return and receiver response | Do not infer far-end SNR from laser power alone |
| Receiver noise | Dark noise, NEP/input noise and bandwidth for each channel | Include APD gain, TIA and ADC noise |
| Far-end SNR | Anti-Stokes ____; Stokes ____ | Check worst-case temperature and loss |
| Temperature-error budget | Photon noise, electronics, gain and attenuation errors | Use error propagation or Monte Carlo |
7 Dual-band receiver and acquisition
Both approximately 870 and 943 nm lie within silicon detector response. A Si APD can improve weak-return sensitivity, but adds bias, temperature compensation, excess-noise and high-voltage design constraints. Select against measured spectral response, dark current, bandwidth, linearity and temperature coefficient.
The receiver should provide:
- Adequate responsivity in each filter passband and measurable channel-response difference.
- Low dark noise and input-referred noise.
- Bandwidth matched to 5-20 ns pulses without unnecessary noise.
- Controlled overload recovery against strong near-end returns and pump leakage.
- Stable APD bias and temperature monitoring.
- ESD, reverse-overvoltage and transient protection.
Keep the two TIA channels structurally similar, with matched supply and thermal conditions, to minimise gain and phase differences. Attenuate pump leakage optically before the detector; software cannot recover a saturated pulse. Provide dark, known-power, test-pulse and temperature-monitor nodes for board calibration and production self-test.
| Detector requirement | Design/selection target | Validation |
|---|---|---|
| Spectral response | Adequate responsivity within each selected passband | Calibrate channel-response difference |
| Noise | Low dark current and input-referred noise | Dark noise spectrum, time noise and temperature drift |
| Bandwidth | Matched to 5-20 ns pulses without excess noise | Frequency, rise-time and pulse-response test |
| Linearity and overload | Recover from strong near returns and residual pump | Overload, recovery and input-power sweep |
| Bias and temperature | Stable APD bias with usable temperature compensation | Gain versus bias and case temperature |
| Reliability | Protect against reverse overvoltage, ESD and transients | Power sequencing and input-protection review |
Acquisition recommendations:
| Item | Recommendation | Reason |
|---|---|---|
| Synchronisation | Two synchronous ADCs or two ADCs sharing one clock | Measure and compensate fixed channel skew |
| Sample rate | Evaluate 250-500 MS/s for 10 ns pulses | Analog bandwidth and pulse shape also set resolution |
| Resolution | Evaluate 12-14 bit ADCs | ENOB, full scale and front-end noise set usable range |
| Averaging | On-FPGA synchronous accumulation | Prevent overflow and record the actual count |
| Time base | Common trigger, timestamps and calibrated delay | Prevent ratio errors from channel misalignment |
| Data logging | Raw waveforms, metadata and calibration version | Enable offline recomputation and regression |
Do not use one analog gain that saturates the near end and pushes the far end into ADC noise. Evaluate dual gain, adjustable TIA, segmented ranges or multiple acquisitions, while accounting for stitching errors.
8 Temperature inversion and calibration
Processing sequence:
- Validate trigger, missing frames, saturation and abnormal shots.
- Remove dark-signal mean and slow baseline while retaining noise statistics.
- Calibrate t0 and map samples to distance.
- Synchronously average, filter and bin the anti-Stokes and Stokes curves.
- Correct filter transmission, detector responsivity and TIA gain.
- Compute R(z) and flag non-positive or low-SNR regions.
- Correct differential attenuation and splice steps.
- Invert temperature and detect implausible jumps.
- Report temperature, SNR, saturation, confidence and timestamp.
Calibration must be tied to the actual fibre configuration. A single set of constants should not be reused across different cable, splice or connector configurations.
| Calibration item | Method | Output |
|---|---|---|
| Temperature constant | Stable reference fibre at two or more temperatures | C, gamma, reference uncertainty |
| Differential attenuation | Uniform-temperature or calibrated two-ended/reference method | delta-alpha or distance-dependent correction |
| Splice correction | Stable zones either side of a splice | Correction position and amplitude |
| Time zero | Known length or reflection/test pulse | t0, ng and range bias |
| Receiver channels | Known optical power into each band | Responsivity, gain ratio, linear range and drift |
| Pre-run check | Warm-up, dark sample and reference-zone verification | Self-test record and drift threshold |
Quality flags are mandatory. Mark low SNR, saturation, bend or splice artefacts, laser drift and expired calibration instead of reporting an unqualified exact temperature. After changing fibre, connector, splice or calibration, require a new verification.
9 Validation and acceptance
Run device characterisation, optical bench tests, short-fibre heated zones, range extension, long-run stability and final design verification in that order. Freeze acceptance against the target market and IEC 61757-2-2 or other applicable standards after checking their current editions.
| Stage | Test | Record | Exit condition |
|---|---|---|---|
| A | Laser spectrum/pulse/energy; APD/TIA noise; filter transmission and rejection | Raw waveform, spectrum, uncertainty and ambient conditions | Repeatable and traceable |
| B | 905 nm rejection and separation of 870/943 nm | Power meter/spectrometer, overload test | Two channels readable; pump leakage controlled |
| C | 10-100 m fibre with known heated zone | Reference thermometers and thermal cycles | Correct location, temperature trend and repeatable calibration |
| D | Extend fibre to 0.5 and 1 km | Far-end SNR, error and update time | Stage target or quantified link shortfall |
| E | Long-run, case-temperature, supply, bend and splice variation | Continuous data and events | Drift and faults inside the frozen threshold |
| F | Design verification against project requirements | Full report and issue closure | Product-spec evidence |
At minimum record wavelength stability, pulse timing, in-fibre energy, two-channel response, distance/spatial resolution, temperature error and repeatability, differential attenuation, far-end capability, drift, and fault/safety behaviour. Preserve raw dual-channel waveforms, sampling rate, averaging count, calibration version, fibre serial/length and ambient temperature. Publish statistics beyond the smoothed curve: mean error, maximum absolute error, standard deviation, valid-point fraction, far-end SNR and update period.
| Test ID | Test | Required record |
|---|---|---|
| T-01 | Spectrum and wavelength stability | Temperature, current, pulse width, repetition, peak wavelength and FWHM |
| T-02 | Pulse timing | FWHM, edges, overshoot, jitter and repeatability |
| T-03 | In-fibre pulse energy | Power/energy, optical losses and time series |
| T-04 | Dual-channel receiver | Responsivity, dark noise, bandwidth, linearity and recovery |
| T-05 | Distance/spatial resolution | Fibre length, heated-zone length and sample spacing |
| T-06 | Temperature error/repeatability | Reference temperature, readings, mean, bias and standard deviation |
| T-07 | Differential loss/splices | Splice positions, channel steps and error before/after correction |
| T-08 | Far-end capability | Length, anti-Stokes/Stokes SNR, averaging count and update time |
| T-09 | Drift and supply variation | Case temperature, supply and temperature-reading drift |
| T-10 | Fault and safety | Open fibre, over-temperature, lost trigger, detector overload and interlock |
10 Risks and controls
| Risk | Effect | Control | Priority |
|---|---|---|---|
| Unsuitable short-wavelength fibre | Weak anti-Stokes at far end; bend artefacts | Characterise 870/905/943 nm and bend/splice effects first | High |
| 905 nm leakage into receiver | APD/TIA/ADC saturation and long dead zone | Optical rejection, overload recovery test, receiver protection | High |
| Spectral/gain/timing mismatch | Bias or false fast-temperature error | Matched layout, response correction and skew calibration | High |
| Uncompensated differential loss | Distance-dependent or local temperature error | Dedicated fibre calibration, splice records and uniform-fibre tests | High |
| Laser pulse drift or heating | Amplitude and timing drift | Per-pulse monitor, warm-up, thermal control and alarms | Medium-high |
| APD bias/temperature instability | Gain/noise drift or device damage | Low-noise bias, temperature compensation and overvoltage protection | Medium-high |
| Driver EMI | TIA baseline disturbance and false triggers | Physical partitioning, grounding, shielding and clock layout | Medium-high |
| Wrong averaging/update-time choice | Noisy or slow measurement | Freeze noise versus update profiles | Medium |
| Invisible 905 nm exposure | Eye hazard without visual warning | Product risk analysis, enclosure, interlocks, labels and training | High |
| Calibration not updated after cable change | Incorrect temperature | Bind calibration to fibre configuration and require revalidation | High |
Close the four highest-risk issues first: pump leakage, differential attenuation, short-band fibre suitability and laser/detector overstress. Only then start long environmental validation.
11 Prototype BOM and interfaces
| Module | Main parts/materials | Interface and selection constraint | First output |
|---|---|---|---|
| Source | 905 nm laser, thermal structure, pulsed driver | Confirm peak current, duty cycle and thermal design from the datasheet | Source characterisation report |
| Monitor path | Photodiode and sampling coupler | Stable split ratio without material main-path loss | Shot-to-shot energy log |
| Common path | Broadband circulator/coupler and fibre adapters | Cover pump and both Raman bands; measurable loss/return loss | Optical loss table |
| Spectral split | 905 nm blocker, 870/943 nm filters and dichroics | Measure real incidence angle, passbands, rejection and drift | Transmission/isolation report |
| Sensing fibre | Candidate fibre, connectors and splice protection | Define core, NA, mode, attenuation and temperature rating | Fibre selection report |
| Detectors | Two Si APDs or suitable Si detectors | Independent bias, temperature monitoring, dynamic range and bandwidth | Noise/responsivity table |
| Analog front end | Two TIAs, limiting/protection and gain control | Low noise, matched bandwidth and controlled recovery | Analog-chain report |
| Acquisition | Dual high-speed ADC, FPGA/MCU and common clock | Synchronous capture, throughput, accumulation and trigger I/O | Timing/data-integrity report |
| Software | Acquisition, calibration, inversion, diagnostics and UI | Version control and raw-data export | Algorithm regression record |
| Test equipment | Spectrometer, power/energy meters, oscilloscope, thermometers and chamber | Calibrated for bandwidth and range | Equipment and calibration list |
Required interfaces include laser enable, pulse width, repetition rate, external trigger and fault/interlock; current, voltage, case temperature, optical monitor, APD bias/temperature and ADC overload; two raw waveforms, timestamps, calibration and quality flags; and maintenance functions for dark capture, test injection, version and event logs.
12 Milestones and decision gates
| Stage | Work package | Deliverable | Entry to next stage |
|---|---|---|---|
| 0 Requirements | Freeze range, temperature, error, resolution, update and cable | SRS and risk list | Every critical metric is measurable and owned |
| 1 Feasibility | Measure candidate fibre at 870/905/943 nm, filters and detectors | Fibre report and first link budget | Short-run dual-band returns and pump rejection |
| 2 Module validation | Test source, receiver and acquisition separately | Subsystem reports | Pulse and noise inside bench target |
| 3 Prototype integration | Common-path optics, timing, dual acquisition and temperature software | V0.1 prototype, raw data and algorithm version | Known zone locates and inverts repeatably |
| 4 Range/accuracy | Longer fibre, steps, differential loss and stability | Verification report | Frozen target met or quantified gap |
| 5 Engineering | Structure, thermal, EMC, reliability, safety and production test | Engineering design and BOM | Risks closed enough to start productisation |
Gate 1: if the 870 nm channel lacks SNR on the target fibre, do not simply raise 905 nm peak power. Revisit fibre type, loss, filtering, receiver bandwidth, averaging and pump band.
Gate 2: if differential attenuation cannot be compensated on known zones and long fibre, do not use the data for field temperature alarms.
13 Safety, reliability and controlled documents
905 nm is nearly invisible and may not be noticed. Risk assessment must include accessible emission, pulse energy, repetition rate, beam geometry and fault states. IEC 60825-1:2014 remains an important classification reference; verify the current edition and market requirements before design freeze or certification.
- Enclose the optical path where practical and provide interlocks or a controlled no-emission procedure for open ports and maintenance.
- Provide visible safety labels, wavelength/output information and maintenance instructions.
- Use current limit, over-temperature shutdown, trigger fault shutdown and fail-off power sequencing.
- Provide safe discharge for APD bias and inspection procedures that never look directly into a fibre end.
Before productisation, validate operating and storage temperature, thermal cycling, vibration, transport, supply disturbance, long-run operation, connector mating, fibre bend and recalibration after repair. Freeze levels for the target power, industrial, infrastructure or laboratory environment rather than copying an arbitrary test class.
Controlled records should include the SRS and trace matrix, optical link budget, fibre selection, laser/driver and receiver test reports, temperature algorithm and calibration procedure, raw-data archive rules, software versions, test specification, risk/DFMEA, laser-safety assessment and user maintenance instructions.
14 Conclusion and next actions
The 905 nm Raman DTS can be engineered as pulsed pump, dual-band Raman receive, time-domain location and temperature inversion. It cannot be reduced to attaching a rangefinder to a telecom fibre. Short-band fibre behaviour, weak anti-Stokes return, 905 nm isolation, differential attenuation and calibration are first-round blockers.
- Freeze the application, range, temperature, resolution, update time, installed fibre and environment.
- Test 2-3 candidate fibres at 870, 905 and 943 nm, including bend and splice effects.
- Build an optical bench and prove stable dual-band returns and pump isolation before selecting the final laser condition.
- Determine pulse width, in-fibre energy, repetition rate and driver from the link budget, not from peak power alone.
- Prove temperature inversion on a short fibre and known zones before extending to 0.5-1 km.
- Freeze acceptance methods and report measurement conditions, statistics and uncertainty.
Appendix A: formulas
| Item | Formula | Note |
|---|---|---|
| Pump wavenumber | v0 = 10^7 / lambda0(nm) | About 11049.7 cm-1 at 905 nm |
| Raman wavelengths | anti-Stokes and Stokes wavenumbers are v0 +/- delta-v | About 870.3/942.5 nm for 440 cm-1, then tune filters to measured spectra |
| Resolution | delta-z approximately ctau/(2ng) | Measured resolution also depends on bandwidth and processing |
| Distance | z = c(t-t0)/(2ng) | Calibrate t0 |
| Repetition limit | f_rep < c/(2ngL) | About 20.4 kHz at 5 km and ng 1.468 |
| Pulse energy | Ep approximately Ppeak*tau | Integrate the real pulse |
| Average power | Pavg = Ep*f_rep | Thermal, power and safety input |
| Raman temperature constant | gamma = hcdelta-v/kB | About 633 K for 440 cm-1 |
| Temperature model | T(z) = gamma/[C-delta-alpha*z-ln(R(z))] | Keep sign convention consistent with calibration |
Appendix B: requirement trace template
| ID | Requirement | Section | Validation |
|---|---|---|---|
| SYS-001 | Independently receive both Raman bands | 3, 6, 7 | T-04 |
| LD-001 | Repeatable 905 nm optical pulse | 5 | T-01/T-02 |
| OPT-001 | Pump leakage does not cause sustained saturation | 6, 7 | T-04 |
| FIB-001 | Three-band attenuation is measured and recorded | 6 | Fibre characterisation |
| ALG-001 | Dual-channel and differential-loss correction works | 8 | T-06/T-07 |
| DIST-001 | Distance mapping meets the requirement | 2, 8, 9 | T-05 |
| TEMP-001 | Temperature error/repeatability meets the frozen target | 4, 8, 9 | T-06/T-08 |
| SAFE-001 | Laser, electrical and fault shutdown verified | 13 | T-10 |
Source documents and standards
Download the complete Chinese engineering report. The source document preserves the full tables, version record and engineering boundaries.
Relevant references include IEC 61757-2-2:2016 for distributed temperature sensing, Hausner et al. on calibration of single-ended Raman DTS, ITU-T G.652 for single-mode fibre, IEC 60825-1 for laser safety and public DTS primers on anti-Stokes/Stokes calibration. Verify current editions, full standard text and certification requirements for the target market.
Product reference: LZFL905-70W-105/125. The exact 62.5/125 um option, power reference plane and connector interface require written confirmation.
Products
Related products
Fiber-coupled package
LZFL905-70W-105/125
905 nm multi-mode pulsed fiber source, 50 W typ in optical table
Calculators
Calculate with this datasheet
Each calculator below opens with this part's datasheet numbers already filled in, in the units the fields ask for.
Pulse Peak Power
Instantaneous peak power from pulse energy and pulse width.
Peak power 50 W · Pulse energy 0.0025 mJ · Pulse width 50 ns Laser FundamentalsAverage Power
Continuous average power from repetition rate and energy per pulse.
Energy per pulse 0.0025 mJ · Repetition rate 10 kHz Laser FundamentalsDuty Cycle & Thermal Load
Fraction of time the laser is on, from pulse width and repetition rate.
Duty cycle 0.05 % · Pulse width 50 ns · Repetition rate 10 kHz Laser FundamentalsPulse Spatial Length
Physical distance an ultrashort pulse occupies in space.
Fiber & LasersFiber Coupling Efficiency
Fiber coupling loss caused by transverse offset of a Gaussian beam.
Fiber & LasersV-Number & Modes
Normalized frequency V of a fiber, and the single-mode versus multimode cutoff condition.
Normalized frequency V 80.1887 · Core radius a 52.5 μm · Numerical aperture NA 0.22 · Operating wavelength λ 905 nmRelated reading
785 nm multimode homogenized-fiber laser: uniform illumination, total-station optics and validation limits
A practical guide to a 785 nm, 90 mW, 200/340 um homogenized-fiber laser, covering illumination uniformity, total-station auxiliary optics, power budgeting and measurement.
Selection guide905 nm vs 1535 nm vs 1550 nm: choosing a wavelength for ranging and LiDAR
The wavelength decision sets your detector, your eye-safety class, your maximum usable energy and your weather performance. Here is how the three bands actually compare for ranging work.
Selection guideAI laser rangefinding glasses: implementation plan and BOM
An implementation blueprint for TO38 905 nm pulsed ranging, APD/TIA/TDC receive electronics, AI flag detection, HUD, power, calibration, production test and laser safety in wearable golf glasses.